72
Preparation
● Turn on the balance:  press e
> The Sartorius logo is displayed; the self-test is performed
● Select the Density application in the Setup menu:  press s
● Select the
Application parameters:  press the q soft key 2 x, 
then the 
O soft key once
● Select
Application 1 (basic settings): press the O soft key
● Select
Density: press the Q or q soft key, repeatedly, if necessary 
● Confirm
Density: press the O soft key
Density Method Density of liquid
ο Buoyancy
Displacement
Pycnometer
Liquid causing ο Water
buoyancy Ethanol
User-definable
Decimals for disp. of None
vol./density 1 decimal place
ο 2 decimal places
3 decimal places
4 decimal places
Printout ο None
All data
ο = factory setting
see also the “Application Parameters (Overview)” in the chapter entitled 
“Configuring the Balance”
● Save settings and exit the Setup menu:  press the 
oo soft key
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
● Press the
CAL soft key
> See “Calibration/Adjustment” 
for further instructions
Toggling to the Next Application
● Press d
> See the section on the 
corresponding application program 
for further instructions
Setup (Setting Parameters)
● Press s
> See “Configuring the Balance” for 
further instructions
Turning Off the Balance
● Press e
> The balance shuts off
> The display goes blank, then 
OFF or Standby is displayed with 
backlighting