Turning the Display On and Off
Press the ON/OFF button
(11)
to turn the display on or off. You
can also turn it on by pressing the tare control
(10)
.
After the balance is connected to line power, the weight display
will go out whenever you turn off the power with the ON/OFF
key. The electronic circuits remain energized (indicated by
STANDBY).
Self-Test
After the power is turned on, a test of all essential electronic cir-
cuits is automatically run.
The display readout "0.000 mg" at the end of the self-test indi-
cated that it has been successfully completed.
Weighing
At this point, place your sample on the pan
(1)
to determine the
weight. Read off the weight in the display
(12)
once the weight
unit symbol (in this case "g") appears. The weight readout is
stable a few seconds afterwards, as a rule.
Taring
lf you wish to use a Container of if the weight display does not
indicate "0.000" mg, press the tare control to zero the display.
This means the balance is ready to operate without requiring
warmup the next time you turn it on.