Troubleshooting Guide
Problem ... Causes ... Remedy
No segments appear in the weight
display
(12)
- No line current available
- The power supply is not plugged
in
- Check power supply
- Plug in power supply
Weight display only shows
"L"
or
"CH 2"
- The weighing pan
(1)
or the pan
support
(2)
is not in place
- Position the pan and pan support
Weight display shows
"H"
- Sample exceeds weighing range - Unload the weighing pan
The weight readout changes
constantly or the special message
"BUSY"
does not go out in the
weight display
- The air bubble of the level
indicator is not within the circle
- Unstable ambient conditions
- Too much vibration or balance
exposed to drafts
- The doors of the weighing
chamber are not completely
closed
- Sample does not have a stable
weight
- level balance
- Set up balance in another area
- Access the menu to adjust the
proper code for the particular
type of weighing environment
- Close the doors of the weighing
chamber
The weight readout is obviously
wrong
- The balance has not been
calibrated
- The balance has not been tared
before weighing)
- The air bubble of the level
indicator is not with in the circle
- Calibrate the balance
- Tare before weighing
- Level the balance