Error Codes
Error codes are shown on the main display. Err codes are shown continuously; Inf messages are shown for 2 seconds, after
which the program returns automatically to the weighing mode.
Display Cause Solution
Err 101 - 104 Key is stuck Release key or
Key pressed at power on contact your local Service Center
Err 320 Program memory defective Contact your local Service Center
Err 335 Verified weighing platform not compatible Connect a compatible weighing platform
with the connected terminal
Err 340 Operating parameter memory (EEPROM) Turn the scale off and then on again.
defective If Err 340 is still displayed, contact
your local Service Center
Err 341 Flash lost data Turn the scale off and then on again
If Err 340 is still displayed, contact
your local Service Center
Err 343 Data lost from the memory module for Contact your local Service Center
transaction numbers in external alibi memory
Inf 01 Data output not compatible with output Change the menu settings
format
Inf 02 Calibration/adjustment condition Calibrate only when zero is displayed.
not met; for example, not tared or Press ) to tare
there is a load on the weighing pan Unload the scale
Inf 03 Adjustment could not Allow the scale to warm up again and then
be completed within a certain time repeat the adjustment process
Inf 06 Built-in calibration weight defective Contact your local Service Center

Table of Contents

Other manuals for Sartorius MIS2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MIS2 and is the answer not in the manual?

Sartorius MIS2 Specifications

General IconGeneral
BrandSartorius
ModelMIS2
CategoryAccessories
LanguageEnglish

Related product manuals