USER MANUAL (Instruction for Use) SW 2.9
Installation
Version 2.9.7 dated 2020-12-23 Page 45 of 75
_
5.3.4 Transportation Damages
If any damages are discovered during or after delivery, please proceed as follows:
In case of outer damage, transporting agent has to be informed, to check damage and
record it on the delivery note.
In any case of damage call SCHWIND eye-tech-solutions or responsible representative
immediately to prepare a report with cost estimation for the insurance.
If possible, prepare photos of the damaged packing.
Store packing material until all questions with transport agent and insurance company
are cleared.
IMPORTANT NOTES
Transportation damages are the responsibility of the transporting agent.
Upon receipt of goods, please note any and all damage, regardless of how minor,
and accept delivery only under written conditions!
5.4 Room and Installation Requirements
Refer to the technical data listed in chapter 10 Technical Data.
CAUTION
When installing the medical system, take care of allowing sufficient space
between the system and other medical devices which may be situated in the same
treatment room.
Electrostatic processes can damage electronic assemblies.
5.5 Device Installation / Start-Up
Unpacking, connection of the system components to the Carriazo-Pendular, initial
installation/start-up and instruction in operating the system (according to MPG) will be performed
by a trained service technician/engineer of SCHWIND eye-tech solutions or authorized
Representative.
The service technician inspects the room for necessary installations (dimensions, electrical
connection, security regulations and air-condition).
He confirms the correspondence in the inspection protocol and connects the system to the power
supply. After a system check and necessary adjustments, he performs operational tests.
IMPORTANT NOTES
The customer is not allowed to perform the electrical connection (main supply and
interlock) of the device or perform the initial installation/start-up.