6.3 Software Operation
6.3.1 Standard System Messages
After attaching the console to voltage supply, the standard system messages appear. Here are all
messages (except error messages), which appear during a system operation.
The console also can be operated without mains connection. For this, the message Mains power
off“ must be confirmed by the operator (button edit/store).
Standard Messages Explanation / Action
Initialisation
Please wait…
System is booting, please wait. The system will create an OK-
melody when ready.
“Annual service is overdue”
“technical check required!”
System has to be checked technically.
Please inform service department. This information can only be
reset by the manufacturer and has to be acknowledged with the
keyboard.
Standby
“carriazo-pendular”
“microkeratome”
VAC-Pedal can be used to test the vacuum, once after switching
on the console, or to start the operation:
“Low battery alert!”
“System will shut down!”
The accumulator is empty and the Carriazo-Pendular will switch
off automatically.
Check the Accu charging fuse (f4) on the rear side.
Check the mains cable = connection to net.
“Vacuum off!”
“Motor test mode: reverse”
Message appears, if vacuum is off and the reverse pedal is
pressed.
Vacuum test
“Vacuum test! Min:
XXXmmHg
XXXmmHg
Automatic test of the pump and the valve, once after switching
on the console. The vacuum has to reach the specified op-
vacuum pressure within a defined time limit and then the
pressure has to be released within a defined time limit.
Operations can only be performed once this test has been
completed successfully.
“Reduce vac! Max:
XXXmmHg”
XXXmmHg
Vacuum out of range! Reduce vacuum with vacuum regulator.
“Vacuum OK. Ready for use!”
System ready for treatment, you will hear an OK-melody.
Operation
“Creating vacuum…”
XXX mmHg
Movement of the head can only be started using the foot pedals
once the stable OP vacuum has been reached.