Page 44
Page 45
9 Maintaining your STATIM
9.7 Maintaining Fluid Levels
1. The water reservoir level is continually monitored by your STATIM. If the reservoir is low, a red X will
appear on the water icon
of the cycle select screen. Press the icon to go to the next screen to
confirm that it is a water level issue and not a water quality issue.
2. If the reservoir is low, a
will appear next to WATER LEVEL.
3. To fill the reservoir, use steam-processed distilled water or VistaPure deionized water containing less
than 5 ppm total dissolved solids (having conductivity of less than 10 μS / cm). Remove the cap from
the top of the unit and fill the reservoir. We recommend using a funnel to minimize spills. Each time
you refill the reservoir, empty the waste bottle and refill with water to the MIN line. Empty the waste
bottle often to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-free
disinfectant, prepared according to the manufacturer’s instructions, may be added to the waste bottle
to remedy this situation).
9.8 Reading Water Quality
1. The water quality is continually monitored by your STATIM to ensure that the optimal quality of water
is being used. If the water quality is beyond the tolerances specified above, a red X will appear on the
water icon
of the cycle select screen. Press the icon to go to the next screen to confirm that it is a
water quality issue and not a water level issue.
2. If the water quality not suitable,
will appear next to micro S. and parts per million values.
3. Using the drain tube (see Section Priming the Pump, Figure 6) empty the contents of the reservoir
into the water container and replace with steam-process distilled water or VistaPure deionized water
containing less than 5 ppm total dissolved solids (having conductivity of less than 10 μS / cm).
9.9 Using On-Screen Instructions
The STATIM G4 on-screen help instruction can be viewed by following these steps:
1.
2. Scroll to
Instructions
and select.
3. Choose the Instructions you would like to view and select.