9 Maintaining your STATIM
Technician
Once a year
Cassette • Check the tray, lid and seal for damage. Replace if necessary.
Biological Filter • Inspect the biological filter for moisture.
Solenoid Valve • Inspect the valve and clean if dirty. Replace the plunger if defective.
Pump • Clean the filters, replace if dirty.
Check Valve
• Remove the exhaust tube from the back of the unit during the air drying
phase. Check for air coming from the fitting.
• Remove the air compressor tube from the check valve inlet while running
a cycle. Make sure no steam is leaking from the valve. Replace if there are
any leaks.
Water Reservoir
• Check the reservoir for dirt. Clean and rinse with Steam distilled water or
VistaPure deionized water if necessary.
Calibration • Calibrate the unit.