DC.R340TT.MAN.00101User Manual SDT340TANKTEST EN
16/44
Node or memory location occupied
Node or memory location selected
Remaining nodes or memory locations
Go to the next step with the F3 key.
Confirm the selected choice with ENTER
The screen summarizes the acquisition time settings. It is only possible to adjust the acquisition time
of the sensors in "TEST" mode. The references associated with the respective background noises of the
sensors, on the day of the test, before application of the vacuum are determined on the basis of an
acquisition time fixed at 6 seconds, per sensor.
The amplification is preset to +90 dB and will adjust automatically. Adjust the audio volume using the
Left Arrow and Right Arrow keys.
On each sensor, the RMS value is automatically updated 2 times per second.