as well as in the TankTest Reporter software used for the edition of the reports, after having exported
the data on your computer.
❶ Identification of the memory location
(Tank4/Ref) as well as the sensor
considered (TTB2•= airborne phase)
❷ Reminder of the sensor configuration
(blocked on A=90 dB with an acquisition
time of 6 seconds)
❸ Real time value of the measurement
❹ Trend observed in the last 30 seconds
❺ Return to the previous screen with the
F1 key
❻ Switch to red sensor (i.e. TTR2 •= liquid
phase)
It is necessary to take successively, a reference measurement with the TTB2 sensor
as well as a reference measurement with the TTR2 sensor (in the liquid), to continue
the guided mode. The latter are useful for characterizing the "ambient" background
noise in the tank, in the airborne phase and in the liquid phase.
Press the Enter key to initiate signal collection on the corresponding sensor. A loading screen will
inform about the progress of the measurement. The trend is proposed on the screen, in addition to
the audio rendering.