84 •
The current setting is displayed.
5. Select the desired setting:
–On
–Off
6. Confirm your selection.
You exit the menu automatically.
7. If you also want to activate beeps for the second
function, repeat the process.
Setting filtering (Fil) Filtering (Fil) allows you to reduce interference (e.g. as a
result of patient movement) when determining weight.
1. Select the item "Fil" from the menu.
2. Confirm your selection.
The current setting is displayed.
3. Select a filtering level.
0: no filtering
1: moderate filtering
2: high filtering
4. Confirm your selection.
You exit the menu automatically.
Switching unit of weight
(Unit)
On non-calibrated scales, you can select the unit (Unit)
in which you want weight to be displayed.
CAUTION!
Patient hazard
In order to avoid misinterpretations, test results
for medical use must be displayed and used in
SI units only. Some devices offer the ability to
display test results in other units. This is only an
additional function.
Use the results exclusively in SI units.
The use of measurement results in non-SI
units is the sole responsibility of the user.
1. Select the item "Unit" from the menu.
2. Confirm your selection.
The current setting is displayed.
Select the unit in which you want weight to be
displayed:
9r: kilograms (kg)