Troubleshooting • 95
English
Only the "SYS" and "Lrn"
items are visible in the
"rF" menu.
The wireless module is activated and no
wireless group is set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 88).
The items "APrt" and
"Time" are not visible in
the "rF" menu.
No wireless printer registered to the wireless
group.
- Register the wireless printer to the wireless
group via the "Lrn" menu item (see "Setting
up a wireless group (Lrn)" on page 88).
After calling up the
menu, the "rF" item is
not displayed.
The scale’s wireless module is defective.
- Inform seca service.
The display "Er:X:11"
appears.
The scale has too high a load or too high a load in
one corner.
- Take the load off the scale or distribute the
weight more evenly.
- Restart the scale.
The display "Er:X:12"
appears.
The scale has been switched on with too high a
load.
- Remove the load from the scale.
- Restart the scale.
The display "Er:X:16"
appears.
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
The Enter key (send/
print) is pressed and the
display "Er:X:71"
appears.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 88).
The Enter key (send/
print) is pressed and the
display "Er:X:72"
appears.
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 88).