EasyManua.ls Logo

Seca 777 - Technical Data; General Technical Specifications; Weighing Data Details

Seca 777
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data • 29
English
17-10-07-651-100a/07-2019S
10.TECHNICAL DATA
10.1 General technical data
10.2 Weighing data
General technical data
Dimensions
Depth
Width
Height (measuring rod not extended)
576 mm
410 mm
1356 mm
Weight of measuring station approx. 12.7 kg
Ambient conditions, operation
Temperature
Air pressure
Humidity
+10 °C to +40 °C / +50 °F to 104 °F
700 1060 hPa
30 % 80 %, no condensation
Ambient conditions, storage
Temperature
Air pressure
Humidity
-10 °C to +65 °C / +14 °F to 149 °F
700 1060 hPa
0 % 95 %, no condensation
Ambient conditions, transport
Temperature
Air pressure
Humidity
-10 °C to +65 °C / +14 °F to 149 °F
700 1060 hPa
0 % 95 %, no condensation
Power supply
Batteries
- Supply voltage
- Battery type
Plug-in power supply unit (optional accessory)
- Supply voltage
- Maximum current consumption
- Insulated device, protection class II (EN 60601-1)
9 V
6 x type AA, 1.5 V
12 V
typically 500 mA
Supply voltage 100 V 240 V
Power supply frequency 50 Hz 60 Hz
Current consumption approx. 37 mA
Maximum runtime in battery operation (USB interface not
active) approx. 32 hours
Metrology data, measuring rod
Measuring range 1
Measuring range 2
Graduations
Accuracy
10 cm 122 cm (4 inch 48 inch)
122 cm 230 cm (48 inch 90.5 inch)
1 mm (1/8 inch)
± 5 mm
Medical device in accordance with Directive 93/42/EEC Class I with measuring function
EN 60601-1: Medical electrical device, Type B
Type of protection IP20
Duty cycle Continuous duty
Weighing data
Maximum capacity 250 kg/550 lbs
Minimum capacity 0.2 kg/0.4 lbs

Table of Contents