EasyManua.ls Logo

Sedecal X - Page 47

Sedecal X
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Universal Radiographic System
Service Manual
SM-0524R4
41
8. T urn ON the Collimator Light. Measure the maximum illumination; this
should occur near the field center. Slide the light sensor along the
Table-Top, and locate the point where the illumination drops to a 75% of
the maximum. This point is defined by BRH as lying on the edge of the
Light Field (ref er to Illustration 4-19). All subsequent measurements will
be referenced to this point and to this definition of edge”.
Illustration 4 -19
Light Contrast Calculation
EDGE OF A QUADRANT
LIGHT SENSOR ON THE LEFT EDGE
BRH POINT
(LIGHT DROPS 25%)
LIGHT SENSOR ON THE RIGHT EDGE
9. Measure the illumination at a point 3 mm from the edge of the field toward
the center of the field (Light sensor on the left of edge). (Refer to
Illustration 4-19). Record this as I
1
.
10. Measure the illumination at a point 3 mm from the edge of the field away
from the center of the field (Light sensor on the right of edge). (Refer to
Illustration 4-19). Record this as I
2
.
11. Correct the values of I
1
and I
2
by subtracting from each value the ambient
light level measured in step-7. Now divide the corrected value of I
1
by
I
2.
This ratio should be 4 or more.
12. Repeat the process from step-8. for all quadrants of Light Field.
13. If the deviation of Light Field Contrast Ratio is out acceptance limits, it
is necessary check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar window
are not dirty or discolored.
G The Light Field Intensity level shall be higher of 170 lux.
G The ambient light level shall be low as it affects the accuracy of the
measurements.

Related product manuals