Universal Radiographic System
Service Manual
SM-0524R4
43
SECTION 5 MAINTENANCE
The purpose of this Periodic Maintenance is to guarantee a continue safe
performance of the Unit, to increase serviceability, to reduce costs (down time,
repairs, etc.) and to assure the safety (personal risk).
The following checks and maintenance procedures, at the the suggested
intervals, are the manufacturer’s recommendation for the most effective
Periodic Maintenance schedule for this Unit.
Service tasks here described must be performed exclusively by service
personnel specifically trained on this medical system. This maintenance
service should be performed every twelve (12) months after installation.
5.1 GENERAL CLEANING
NEVER ATTEMPT TO CLEAN OR HANDLE ANY PART OF
THE UNIT WHEN IT IS ON. SWITCH IT OFF AND
DISCONNECT IT FROM THE MAINS BEFORE CLEANING OR
INSPECTING.
GENERAL CLEANING
EXTERNAL SURFACES Clean the equipment frequently, particularly if corroding chemicals are present. Clean
external covers and surfaces, especially parts in contact with patients, with a cloth
moistened in warm water with mild soap. Wipe with a cloth moistened in clean water.
Do not use cleaners or solvents of any kind. Also check painted surfaces for scratching
and touch up as required.
INTERNAL CLEANING DANGER: NEVER ATTEMPT TO CLEAN INTERNALLY THE UNIT. THE UNIT
MUST BE ALWAYS KEPT SHUT UNLESS A SPECIFIC RENEWAL PART
PROCEDURE IS TO BE UNDERTAKEN BY HIGHLY QUALIFIED MAINTENANCE
PERSONNEL.