126
K600000000SE 030804 011018
13.10 Measurements → Tightness test TRGI → New
(when the instrument version so provides)
CONTEXT KEY FUNCTION
Enters the modification mode for the selected parameter.
Goes to the next phase of the tightness test.
Performs pressure zeroing.
Interrupts the current phase.
Repeats the tightness test.
Saves, in the memory selected in the "Select Memory" menu, the data
acquired.
Tightness test has been saved.
Starts printing the paper print-out.
KEY FUNCTION
Activates the context keys shown on the display.
Select line; the selected line is displayed in red.
In edit mode, it is the desired value.
Enters the selected parameter setting.
Returns to the previous screen.
When in modify mode cancels the modification just made.
Volume of the gas line <100 or 100..200 or >200 litre.
O
Waiting time 15 ... 240 minutes.
15/01/14
10:00
TRGI
Configuration
Volume
l
<100
Stabilization
min
15
Test duration
min
15
Print
manual
Duration time of test 15 ... 240 minutes.
Print out test of the result (manual or automatic).