7.0 TECHNICAL SPECIFICATIONS
7.1 Technical features
Power supply: Li-Ion battery pack with internal protection circuit, rechargeable.
Battery charger: External 5Vdc 2A battery charger with female A-type USB connector +
connection to the device with the same serial communication cable
supplied.
Charging time: 5 hours to charge from 0% to 90% (6 hours for 100%). The device can
also be charged by connecting it to the PC, the device must be turned off,
the charging time depends on the output current from the PC and may be
more than 12 hours.
Instrument working time: 12 hours of non-stop operation (excluding printing).
——————————————————————————————————————————————————
Display: 4.3” TFT 272x480 pixels graphic color with backlight
——————————————————————————————————————————————————
Connectivity:
Communication port: USB connector type B.
Bluetooth
®
: Class 1. Communication distance <100 meters (in open field)
——————————————————————————————————————————————————
Autozero: Automatic autozero cycle with the probe not inserted in the chimney.
Dilution: Widens the CO sensor measurement range up to 100.000ppm (10.00%).
Programmable as simple protection of the CO sensor with the intervention
level set by the user.
Pre-set triggering threshold at 1500 ppm.
——————————————————————————————————————————————————
Gas measurement sensors: Up to 6 configurable sensors: electrochemical, NDIR (single cell) and
pellistor.
Programmed fuels: 15 factory pre-set plus 32 user-programmable.
——————————————————————————————————————————————————
Self-diagnosis: All the functions and internal functions are checked and anomalies
signaled.
Temperature measurement: Double K thermocouple input with mini connector (ASTM E 1684-96 ) to
measure differential temperature (supply and return)
Measurement of ambient temp.: Via internal sensor or T2 thermocouple input with remote probe.
——————————————————————————————————————————————————
Printer: Thermal integrated with easy loading paper and sensor for the presence of
paper
Printer powered: By the analyzer batteries.
Printer autonomy: Up to 40 analysis reports with the batteries fully charged.
——————————————————————————————————————————————————
Internal data memory: 2000 complete data analyses, time and name of the customer can be
stored.
User data: 8 programmable user names.
Print header: 6 lines x 24 characters user-customized.
——————————————————————————————————————————————————
Suction pump: 1.0 l/min heads at the flue up to 135mbar.
Measurement of flow: Internal sensor to measure the flow of the pump.
Condensate trap: Outside the instrument.
Line filter: With replaceable cartridge, 99% efficient with 20um particles (inside the
condensate trap).
Smoke: Using an external hand pump; it is possible to enter and print the smoke
index.
Leak test: Gas pipes tested for leaks with separate printout of the result, by means of
the attachment AACKT02, according to UNI 7129 (new systems) and UNI
11137: 2012 (existing systems), with automatic calculation of pipe volume.
Condensing boiler efficiency: Automatic recognition of the condensing boiler, with calculation and
printout of efficiency (>100%) on the LHV (Lower Heating Value) in
accordance with UNI10389-1.
Environmental gases: Measurement and separate printout of the ambient CO values.
Draft test: By using the internal sensor connected to the port P-, resolution 0,1 Pa,
accuracy 0,5 Pa.
Draft test must be carried out only by means of the AACDP02 external
accessory in compliance with UNI 10845 standard.
——————————————————————————————————————————————————
Operating temperature: 23°F to 113° F (-5 °C to +45 °C)
Storage temperature: -4°F to 122°F (-20 °C to +50 °C)
Humidity limit: 20% .. 80% RH