13.10.3 Performing a tightness test for a gas line with volume greater 200 litres.
NOTE: If, while configuring the tightness test the automatic printing mode has been selected, the tightness test is
printed automatically.
Instead, if the manual printing mode has been selected (exemplified case), at the end of the tightness test the results
are displayed and they can be saved and/or printed. In this case proceed as follows.
15/01/14
10:00
TRGI
Stabilization phase
P diff
hPa
0
F1 F2 F3
WARNING
Pressure for testing
Fill the system up to:
150.00 hPa
Press any key
F1
O
15/01/14
10:00
TRGI
Stabilization phase
P diff
hPa
0
O
AUT.NULLUNG
5
Wait for
autozero
completion
then
pressurize
the system.
15/01/14
10:00
TRGI
Stabilization phase
P diff
hPa
0.06
O
automatic
15/01/14
10:00
TRGI
Stabilization phase
hPa
150.45
P diff
hPa
-0.92
ΔP 1 min
00:30:00
wait time
Pressure of the system.
Pressure variation in the last minute.
Waiting time to end the stabilization phase.
15/01/14
10:00
TRGI
Stabilization phase
P diff
hPa
150.06
O
Pressure
for testing:
150.00 hPa
automatic
or
15/01/14
10:15
TRGI
Start test
hPa
141.90
P1
hPa
140.96
P2
hPa
0.44
ΔP
00:60:00
duration time
Pressure at starting point.
Pressure variation in the last minute.
Time to finish tightness test.
Current measured pressure.
The stabilization phase can be interrupted
in any moment.