On site calibration
Sensor CO
15/01/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
not active
Calibrate
Ia
uA
0.17
OK
OK
On site calibration
Sensor CO
15/01/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
active
Calibrate
Ia
uA
0.17
OK
Messages in the 'Status' line:
saving: the instrument is saving the
performed calibration
error: the sensor has NOT been
recalibrated for any of the following
reasons:
- The calibration gas cannot
properly reach the instrument.
- Concentration for the calibration
gas has not been set in the
relevant line ‘Applied gas’.
- The user didn’t allow for the
stabilization time to properly
elapse.
- The sensor could be damaged or
exhausted and must therefore be
replaced.
WARNING
At any time the user can restore the factory calibration in the instrument by setting
the 'Status' line on 'not active'.
When the stabilization time is over, select the 'Calibration' row and activate the function ' ' to store the
new calibration.
OK