LISST-200X User’s Manual LISST-200X Introduction and Tutorial Page 22
If the Start and Stop Conditions are set to Depth, External Mechanical
Switch, or External Digital Input the program will return to checking for
the Start Condition. This will only be true if the Start and Stop
conditions match. For example, Depth Start and Depth Stop. For non-
matching Start and Stop Conditions, such as fixed number of samples
or Time Stop, the program will terminate and the instrument will go into
a low power sleep mode.
To stop a running program, or wake it from low power sleep mode, use
the Stop button on the Terminal window or the button on the tool
bar.
It is good practice to always send a Stop command to the instrument
before attempting any other communication with the instrument.
Step 18:
Downloading
Data
The instrument has now stored data in the on-board memory card.
Pressing the Instrument Query Button , will display the instrument
status, including the number of samples saved. To download the data
to your computer select Download Data From LISST from the LISST
menu or choose the , button from the toolbar. A list of files will
appear:
Choose the files to offload by clicking on them while holding down the
CTRL key. The Shift key can also be used to select a range of files.
Select OK and then choose a location to save the files to.