EasyManuals Logo

Sequoia LISST-200X User Manual

Sequoia LISST-200X
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
LISST-200X User’s Manual LISST-200X Operation Details Page 66
12. Simple Real-Time Data Processing
This step-by-step procedure covers the acquisition, display and storage of processed data in
real time.
STEP
ACTION
RESULT
1
Connect the instrument to the computer and establish
communication in the LISST-SOP200X program. (see
Step-by-Step Procedure: Establishing Communication
with the LISST-200X)
Communication
established.
2
Before starting a Real-Time session the output file options
must be set. This must be done before selecting Collect
Realtime Data. Select the Output tab in the Settings
menu. The window below will appear.
Output file type
selected.
A binary PBN file must be selected before any other file type
will be stored. This is used to display the processed data to
the screen.
The CSV type is a comma delimited file containing all the
processed data.
The RTX file is a comma delimited file containing the raw
data from the data logger.
You must also select the inversion model to be used with
the real time data processing: Spherical, randomly
shaped or both. For details on these selections, see
Appendix G: Spherical and Random Shape Particle
Models.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sequoia LISST-200X and is the answer not in the manual?

Sequoia LISST-200X Specifications

General IconGeneral
BrandSequoia
ModelLISST-200X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals