LISST-200X User’s Manual LISST-200X Operation Details Page 82
TD x or STARTDATA x where x is the start condition data described
below
The TD command sets the start condition data to be used when the GO
command is issued. The start condition data is used with the Start
Condition setting as follows:
• If the Start Condition is Depth Start (option 1) the input will be start
depth in meters.
• If the start condition is set to Time/Date Start (option 2) the input for
TD will be the start date and time.
• If the Start Condition is Delay Start (option 5) the input will be time
delay in minutes.
• The TD setting is ignored for Mechanical Switch Start (option 3) or
Digital Input Start(option 4).
If no value follows command, prompts will be displayed for the value.
input: TD 3 (if Start Condition = 1 (Depth Start))
output: New Start Condition data = 3
Start Condition: Depth Start at 3 meters
input: TD 2016/12/31 23:59:59 (if Start Condition = 2 (Time/Date Start))
output: New Start Condition data = 2016/12/31 23:59:59
Start Condition: Time Start at 2016/12/31 23:59:59
Input: TD 2 (if Start Condition = 5 (Delay Start))
Output: New Start Condition data = 2
Start Condition: Delay Start with 2 minute delay
SP x or STOPDATA x, where x is the stop condition code described below
The SP command sets the stop condition to be used when collecting data.