LISST-200X User’s Manual LISST-200X Operation Details Page 83
The stop condition options are:
1 = Depth Stop
2 = Time/Date Stop
3 = Mechanical Switch Stop
4 = Digital Input Stop
5 = Number of Samples Stop
6 = Maximum memory or Low Battery Stop
If no value follows command, prompts will be displayed for the value. After
setting a stop condition of Depth, Time/Date or Number of Samples, use
STOPDATA to enter an appropriate parameter value.
input: SP 5
output: New Stop Condition Setting: Fixed Number Stop
PD x or STOPDATA x where x is the stop condition data as described
below
The PD command sets the stop condition data to be used when the
collection data. The stop condition data is used with the Stop Condition
settings as follows.
• If the Stop Condition is Depth Stop (option 1) the input will be stop
depth in meters.
• If the stop condition is set to Time/Date Stop (option 2) the input for
PD will be the stop date and time.
• If the Stop Condition is Fixed Number Stop (option 5) the input will
be the number of samples to collect before stopping.
• The PD setting is ignored for Mechanical Switch Stop (option 3),
Digital Input Stop (option 4) or Maximum memory or Low Battery
Stop (option 6).
input: PD 3 (if Stop Condition =1 (Depth Stop))
output: New Stop Condition data = 3
Stop Condition: Depth Stop at 3 meters
input: PD 2016/12/31 23:59:59 (if Stop Condition = 2 (Time/Date Stop))
output: New Stop Condition data = 2016/12/31 23:59:59
Stop Condition: Time Stop at 2016/12/31 23:59:59