LISST-200X User’s Manual LISST-200X Operation Details Page 89
Notes:
The GX command does not store the sample to a datafile.
XR
Transmit raw data during sampling
Syntax:
XR x, where x = 0 through 3
Description:
This command allows data to be transmitted to an external data system
during execution of the GO command, in one of several formats. The
default setting is 0, with data transmission turned off. Setting 1 sends raw
data in a multi-line text format; 2 sends a single line as a NMEA-
compatible data sentence; 3 sends a single line of comma-separated
values. Contact Sequoia Scientific for complete details of these formats.
Example:
input: XR 1
output: Raw data WILL be transmitted when data is stored
Cautions:
Format 2 requires firmware 1.3 or later. Format 3 requires firmware 1.414
or later.
ZS
Collect and transmit background scattering data
Syntax:
ZS
Description:
Acquires 20 averaged sample and displays the result to the screen.
Evaluates the data for quality as a background measurement.
Example:
Input: ZS
Output: {
…. 36 ring values + 24 Aux parameters
}
repeat 20 times, then messages about the quality of the data.
Notes:
The ZS command does not store the sample to a datafile.
ZZ
Go into deep sleep mode (minimum power consumption)
Syntax:
ZZ