16 | Page
INSTALLATION
REQUIRED AMBIENT CONDITIONS
These units are built for use indoors at room temperatures between 15°C and 30°C (59°F and 86°F), at
no greater than 80% Relative Humidity (at 25°C / 77°F The ambient temperature should not change by
2°C (3.6°F) or more during operation.
Operating outside these conditions may affect the incubator temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
Proximity to ovens, autoclaves, and any device that produces significant radiant heat
Heating and cooling vents or other sources of fast-moving air currents
High-traffic areas
Direct sunlight
REQUIRED CLEARANCES
These clearances are required to provide air flows for ventilation and cooling.
4 inches (102 mm) of clearance is required on the sides and back.
2 inches (51 mm) of headspace clearance between the top of the unit and any overhead partitions.
2” (51 mm)
4” (102 mm)
4” (102 mm)
Access Port
2” (51 mm)
4” (102 mm)
Door Swing
23.5” (597 mm)
Power Cord
Gas Connections