31 | Page
OPERATION
Note: From a cold start, the incubator requires 12 hours to come up to and stabilize at temperature
and humidity levels prior to loading samples. Stabilization safeguards samples.
PUT THE INCUBATOR INTO OPERATION
Perform the procedures below after the unit has been installed in a new workplace location. These
prepare the incubator for normal use.
1.
2.
Attach the power cord that came with the unit to the power
inlet receptacle on the side of the incubator.
Plug the power cord into the workspace electrical supply outlet.
3.
Turn on the incubator.
Place the incubator Power Switch in the ON ( I )
position.
4.
Perform the Humidify the Incubator procedure on page 33.
5.
Carry out the Set the Temperature Setpoint procedure on page
34. Set the incubator to your incubation application temperature.
6.
Perform the Start a CO
2
Flow procedure on page 36.
7.
Carry out the Set the CO
2
Setpoint procedure on page 37. Set the
CO
2
setpoint to your application gas concentration.