6 Troubleshooting
6.1 Troubleshooting
74 FPD-2010
■ Baseline drift, meander
Cause Countermeasures
There is a large quantity of impurities in the carrier and
make-up gases.
the Filter (Molecular Sieve Filter, Gas Fil•Replace -
ter or Oxygen trap).
the tubes and joints.•Replace
the gas container, changing to a Gas of
higher purity.
•Replace
Gas leakage Check for gas leakage and correct any piping joints
from which gas is leaking.
Injection port (glass insert) column is contaminated. the glass insert or replace it. (Refer to “3.5
Checking Glass Insert and Injection Port Septum”
• Clean
.)
• Condition the column.
Stray light Determine from where the light is leaking by using a
dark film and stop the stray.
Temperature of each section is unstable. Confirm the temperature of each unit. If it does not sta-
bilizes, contact your local Shimadzu techincal office.
The flow rate is unstable. Confirm the pressure and flow rate of AFC. If it does
not stabilize, contact your local Shimadzu techincal
office.
3.
Zero
Cause Countermeasures
Stray light Determine from where the light is leaking by using a
dark film and stop the stray.
adjustment does not work.
4.
No
Cause Countermeasures
The hydrogen flame goes off. Ignite hydrogen again.
Quenching occurs. Check the analysis conditions. (Refer to “9.1 Quench-
ing”.)
An wrong filter is used. the filter (Refer to “3.3 Comfirming the Filter”.)Confirm
The detector has not been turnd on. on the detector.([DET#], Turn [ON], and [ENTER]
keys.)
Shield ring position is inadequate. Set the shield ring to appropriate position.
peak
5.
Peak is too small (or the sensitivity is too low.)
Cause Countermeasures
The carrier gas is leaking. Check for gas leakage and correct any piping joints
from which gas is leaking.
The split ratio is too large. (in case of split analysis) Reduce the split ratio.
The quantity of the sample injected is too small, or the
sample concentration is too low.
Increase quantity or concentration of injected sample.
The value of range or attenuation is inappropriate. Set the range larger or rise the attenuation.
The capillary column is located at a wrong place. Mount the column at the correct location. (Refer to “3.6
Mounting the Column”.)
An incorrect column is used. Replace the column to the correct one, taking the
polarity into consideration.