8.1
8.8.
79FPD-2010
8 Specifications
8.1Specifications of the FPD-2010
Detection method : Flame photometric method using air-hydrogen flame
Detector : Side-on type Photomultiplier
Range : Five stages form 1 to 10
-4
(at analog linear output)
Attenuation : Five stages from 1 to 2
-4
(at analog wide output)
Dynamic range : S--- 10
3
(Dodecanethiol)
: P--- 10
4
(Tributyl Phosphate)
Minimum detection amount
: S --- 4 pgS/s (Dodecanethiol)
: P --- 0.2 pgP/s (Tributyl Phosphate)
Maximum operating temperature
: 350°C (Continuous use is possible at the room temperature of 25°C or below.)
: 300°C (Continuous use is possible at the room temperature of 30°C or below.)
: 250°C (Continuous use is possible at the room temperature of 40°C or below.)