Display: 3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Inputs: Measurement input µA: (Dead-Stop-) connector for double platinum electrode
Connection sockets: 2 x 4 mm)
Power supply: power supply 90-240 V; 50/60 Hz, power input: 30 VA
RS232-C Interface: RS232-C interface separated galvanically through photocoupler
Daisy Chain function available.
Data bits: adjustable, 7 or 8 Bit (default: 8 Bit)
Stop bit: adjustable, 1 or 2 Bit (default: 1 Bit)
Start bit: static 1 Bit
Parity: adjustable: even / odd / none
Baud rate: adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
Address: adjustable, (0 to 15, default: 01)
RS232-1 for computer, input Daisy Chain
RS232-2 devices of SI Analytics, titrator TitroLine
6000/7000/7500,
- Burettes TITRONIC
500, TITRONIC
110 plus, TITRONIC
universal,
- Balances of the types Mettler, Sartorius,
- Kern, Ohaus (for more, please contact SI Analytics)
- Exit Daisy Chain
USB Interface: 2 x USB-type A and 1 x USB-type B
USB –Typ B (“slave“) for connecting a PC
USB –Typ A (“master“) for connecting:
- USB keyboard
- USB printer
- USB “mouse“ (“mouse“),
- USB data media e.g. USB stick
- USB Hub
Stirrer/pump TMKF: 12V DC out, 500mA
power supply for stirrer TM 235 and KF titration stand TM 235 KF
Housing material: Polypropylene
Front keyboard: polyester coated
Housing dimensions: 15.3 x 45 x 29.6 cm (W x H x D), height incl. interchangeable unit
Weight: ca. 2.3 kg for basic unit
ca. 3.5 kg for complete device incl. interchangeable unit (with empty reagent bottle)
Ambient conditions: Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °