78
Chapter 1 Specifications Titrator TitroLine
®
7500 KF
Status Aug. 23
rd
2011
Interchangeable units
Compatibility: units are compatible to the titrators TitroLine
®
6000, TitroLine
®
7000, TitroLine
®
7500 KF and Piston Burette TITRONIC
®
500
Recognition: automatically through RFID. Recognition of unit size and characteristics of the
Titration- or dosing solution
Valve: volume neutral cone valve made from fluorocarbon polymers (PTFE), TZ 3000
Cylinder: borosilicate glass 3.3 (DURAN
®
)
Hoses: FEP hose set, blue
Bracket for supply bottle: suitable for square glass bottle and misc. reagent bottles
Materials: borosilicate glass DURAN
®
, fluorocarbon polymers (PTFE), stainless steel,
polypropylene,
Dimensions: 15 x 34 x 22.8 cm (W x H x D) incl. reagent bottle
Weight: approx. 1.2 kg for interchangeable unit WA incl. empty reagent bottle
Dosing accuracy: after DIN EN ISO 8655, part 3
Accuracy: 0.15 %
Precision: 0.05 - 0.07 %
(Depending on the used interchangeable unit)
Dosing accuracy of the Titrator TitroLine
®
7500 KF with WA interchangeable units:
Interchangeable.
unit
type No.
Volume
[ml]
Tolerances of the Ø
i
of the glass cylinder
[mm]
Dosage error*
according to
100 % volume
[%]
Reproducibility
[%]
WA 05 5.00 ± 0.005 ± 0.15 0.07
WA 10 10.00 ± 0.005 ± 0.15 0.05
WA 20 20.00 ± 0.005 ± 0.15 0.05
WA 50 50.00 ± 0.005 ± 0.15 0.05