Cleaning procedure:
Switch off the device for the duration of the cleaning operation. If this is not
possible, wear suitable laser safety goggles. These must absorb radiation of the
device’s wavelength effectively.
Glass window: remove dust from the viewing window using a soft, clean brush. If
necessary, also clean the viewing window with a clean, damp, lint-free cloth, and a
mild anti-static lens cleaning fluid.
Plastic window: clean the viewing window only with a clean, damp, lint-free cloth,
and a mild anti-static lens cleaning fluid.
NOTICE
If the inspection window is scratched or damaged (cracked or broken), the lens must be
replaced. Contact SICK Support to arrange this.
If the inspection window is cracked or broken, take the device out of operation
immediately for safety reasons and have it repaired by SICK.
Cleaning the housing
In order to ensure that heat is adequately dissipated from the device, the housing
surface must be kept clean.
Clear the build up of dust on the housing with a soft brush.
8 MAINTENANCE
40
O P E R A T I N G I N S T R U C T I O N S | Inspector85x 8028305//2023-05 | SICK