Technical Information Chapter 3
RFH630
Electrical installation
8014957/YSU6/2016-01-08 © SICK AG · Germany · All rights reserved · Subject to change without notice 25
3.7 Using the CDB620-001 connection module
3.7.1 Wiring overview (one switching input used)
CDB620-001 Connection Module
CMC600 parameter cloning module
(optional)
ONOFF
POWER
S1
0.8 A T
Term CAN
Term 485RS
SGND - GND
422 485
ONOFF
NO
YES
ONOFF
ONOFF
S2 S3
S7S6
S4
CMC
10 11 12 13 14 15 16 17 18
Sens 1
In 1
In 2
U
IN
*
U
IN
*
Sens 2
SGND
SGND
SGND
LEDs
20 21 22 23 24
Res 1
Res 2
Out 1
Out 2
GND
30 31 32 33 34
CAN_H
CAN_L
T+
R+
GND
40 41 42 43 44
CAN_H
CAN_L
T‒/TxD
R‒/RxD
GND
SCANNER
AUX interface
6110 5
1115
5
1
9
6
U
IN
U
IN
GND
GND
Shield
Shield
Shield
Shield
12345678
= an CMC600 is required to provide the additional switching inputs and outputs
DC 10 V ... 30 V
External
reading
clock sensor
(e.g. photo-
electric
switch)
V
S
Out
GND
Result 1
PLC
GND
Result 2
to PC
to
RFH630
Pin
2: RxD
3: TxD
5: GND
Host
TD‒
TD+
RD+
RD‒
TxD
Host
RxD
GND
GND
RS-232
RS-422
V
S
= DC 10 V ... 30 V on terminal U
IN
= U
IN
* after fuse F and switch S1