12
2. Technical data
SITRANS F M
Type MAG 1100 F
Liner
MAG 1100 F (Ceramic) Aluminium oxide Al
2
O
3
(ceramics)
MAG 1100 F (PFA) Reinforced PFA (teflon) (no ATEX)
Electrodes
MAG 1100 F (Ceramic) Platinum with gold / Titanium brazing alloy
MAG 1100 F (PFA) • DN 10 ... 15 (
3
/
8"
... ½
"
): Hastelloy C276
• DN 25 ... 100 (1
"
... 4
"
): Hastelloy C22
Cable entries • Remote installation 2 x M20 or 2 x ½ NPT
• Compact installation
- MAG 5000/MAG 6000: 4 x M20 or 4 x ½
"
NPT
Certificates and approvals
MAG 1100 F (Ceramic) 3A (sensor with Polyamid terminal box ), transmitter not part of the approval
• Ex ATEX approvals for sensor ATEX 2G D sensor EEx d e ia IIB T3 - T6
• Sensor with/without
MAG 5000/6000 FM Class 1 div 2
MAG 1100 F (PFA) 3A (sensor with Polyamid terminal box ), transmitter not part of the approval
FM Class 1 div 2
Conforms to PED – 97/23/EC and CRN (PFA)
Custody Transfer
(MAG 5000/6000 CT) Cold water pattern approval PTB (Germany)
Heat meter pattern approval - OIML R 75 (Denmark)
Hot water pattern approval - PTB (Germany)
Other media than water pattern approval - OIML R 117 (Denmark)
Weld-in adapter
Adapter for welding onto dairy pipe
Tri-Clover, ISO 2037, DIN 11850,
SMS 3008, BS 4825-1
• DN 10, 15, 25, 40, 50, 65 and 80
(
3
/
8
"
, ½
"
, 1
"
, 1½
"
, 2
"
, 2½
"
and 3
"
) PN 40 (600 psi)
• DN 100 (4”) PN 25 (350 psi)
Clamp adapter
Tri-Clamp, ISO 2852, DIN 32676,
SMS 3016, BS 4825-3
• DN 10, 15, 25, 40 and 50
(
3
/
8
"
, ½
"
, 1
"
, 1½
"
, and 2
"
) PN 16 (200 psi)
• DN 65, 80 and 100 (2½
"
, 3
"
and 4
"
) PN 10 (150 psi)
Thread adapter
DIN 11851
• DN 10, 15, 25, and 40
(
3
/
8"
, ½
"
”, 1
"
, and 1½
"
) PN 40 (600 psi)
• DN 50, 65, 80 and 100
(2
"
, 2½
"
, 3
"
and 4
"
) PN 25 (350 psi)
ISO 2853, BS 4825-4
• DN 10, 15, 25, 40, 50, 65 and 80
(
3
/
8
"
, ½
"
, 1
"
, 1½
"
, 2
"
, 2½ and 3
"
) PN 16 (200 psi)
SMS 1145
• DN 25, 40, 50, 65 and 80
(1
"
, 1½
"
, 2
"
, 2½
"
and 3
"
) PN 6 (80 psi)
Design
Material
Adapter Stainless steel AISI 316/Stainless steel AISI 304 (ISO 2852)
Gasket
• MAG 1100 F (Ceramic) FKM/FPM with stainless steel insert (AISI 304) (-20 ... +150 °C (-4 ... +302 °F))
EPDM (-20 ... +150 °C (-4 ... +302 °F))
• MAG 1100 F (PFA) EPDM (-20 ... +150 °C (-4 ... +302 °F))
NBR (-20 ... +100 °C (-4 ... +212 °F))
Note:
When combined sensor and adapter, the operating pressure is the lower rated of the two.