13
SITRANS F M
2. Technical data
Type MAG 3100 MAG 3100 HT (High Temperature) MAG 3100 P
Nominal size DN 15 ... DN 2000 (½
"
... 78
"
) DN 15 ... DN 300 (½
"
... 12
"
) DN 15 ... DN 300 (½
"
... 12
"
)
Measuring principle Electromagnetic induction
Excitation frequency DN 15 ... 65 (½
"
… 2½
"
): DN 15 ... 65 (½
"
… 2½
"
): DN 15 ... 65 (½
"
… 2½
"
):
(Main supply: 12.5 Hz / 15 Hz 12.5 Hz / 15 Hz 12.5 Hz / 7.5 Hz
50 Hz / 60 Hz) DN 80 ... 150 (3
"
... 6
"
): DN 80 ... 150 (3
"
... 6
"
): DN 80 ... 150 (3
"
... 6
"
):
6.25 Hz / 7.5 Hz 6.25 Hz / 7.5 Hz 6.25 Hz / 7.5 Hz
DN 200 ... 1200 (8
"
... 48
"
): DN 200 ... 300 (8
"
... 12
"
): DN 200 ... 300 (8
"
... 12
"
):
3.125 Hz / 3.75 Hz 3.125 Hz / 3.75 Hz 3.125 Hz / 3.75 Hz
DN 1400 ... 2000 (54
"
... 78
"
):
1.5625 Hz / 1.875 Hz
Process connection
Flanges EN 1092-1, raised face EN 1092-1, raised face EN 1092-1, raised face
(EN 1092-1, DIN 2501 & BS 4504 (EN 1092-1, DIN 2501 & BS 4504 (EN 1092-1, DIN 2501 & BS 4504
have the same mating dimensions) have the same mating dimensions) have the same mating dimensions)
DN 65 ... 2000 (2½
"
… 78
"
): DN 15 ... 300 (½
"
… 12
"
): DN 15 ... 50 (½
"
… 2
"
):
PN 6 (87 psi) PN 40 (580 psi) PN 40 (580 psi)
DN 200 ... 2000 (8
"
… 78
"
): DN 65 ... 300 (2½
"
… 12
"
): DN 65 ... 300 (2½
"
… 12
"
):
PN 10 (145 psi) PN 16 (232 psi) PN 16 (232 psi)
DN 65 ... 2000 (2½
"
… 78
"
): DN 200 ... 300 (8
"
… 12
"
): DN 200 ... 300 (8
"
… 12
"
):
PN 16 (232 psi) PN 10 (145 psi) PN 10 (145 psi)
DN 200 ... 600 (8
"
… 24
"
): DN 200 ... 300 (8
"
… 12
"
):
PN 25 (362 psi) PN 25 (362 psi) ANSI B16.5 (~BS 1560), raised face
DN 15 ... 600 (½
"
… 24
"
): ½
"
... 12
"
:
PN 40 (580 psi) ANSI B16.5 (~BS 1560), raised face: Class 150 (20 bar (290 psi))
DN 50 ... 300 (2
"
… 12
"
): ½” ... 12
"
:
PN 63 (913 psi) Class 150 (20 bar (290 psi))
DN 25 ... 300 (1
"
… 12
"
): ½” ... 12
"
:
PN 100 (1450 psi) Class 300 (50 bar (725 psi))
ANSI B16.5 (~BS 1560), raised face
½” ... 24”: Class 150 (20 bar (290 psi)) AS 2129, raised face ½
"
... 12
"
:
½” ... 24”: Class 300 (50 bar (725 psi)) Table E
AWWA C-207, flat face 28
"
... 78
"
: Other flanges and pressure ratings
Class D (10 bar) on request
AS 2129, raised face ½
"
... 48
"
:
Table E
AS 4087, raised face:
PN 16 (DN 50 ... 1200, 16 bar (232 psi))
PN 21 (DN 50 ... 600, 21 bar (304 psi))
PN 35 (DN 50 ... 600, 35 bar (508 psi))
Other flanges and pressure ratings
on request
Rated operation conditions
Ambient temperature
(conditions also
dependent on liner
characteristiques)
• Sensor -40 … +100 °C (-40 … +212 °F) -40 … +100 °C (-40 … +212 °F) -40 … +100 °C (-40 … +212 °F)
• Sensor ATEX -20 … +60 °C (-4 … +140 °F) For medium temperature up
to 150 °C (302 °F)
-20 … +60 °C (-4 … +140 °F)
For medium temperature
150 ... 180 °C (302 ... 356 °F):
-20 … +50 °C (-4 … +122 °F)
Compact transmitter
MAG 5000/6000 -20 … +50 °C (-4 … +122 °F) -20 … +50 °C (-4 … +122 °F) -20 … +50 °C (-4 … +122 °F)
2.3 Sensor MAG 3100, MAG 3100 HT and MAG 3100 P
MAG 3100, MAG 3100 HT and MAG 3100 P