Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
System 13
Page 13 of 34
Medical Solutions
2-
2 System
2.1 Checks
Fig. 1: MAMMOMAT checks
PMP Radiation protection
Check the radiation protection shield for possible damage (1/Fig.1/p.13).
Information regarding the lead equivalent value must be legible.
SIM Swivel arm/PPS tube attachment
Check the swivel arm/PPS tube attachment by pulling on both support rails
(a minimum play of approximately ± 2° is correct). (2/Fig.1/p.13).
SIM Basic table
Check the mounting of the basic table as well as the locking mechanism for the expo-
sure system (3/Fig.1/p.13).
SIM Flying wing (MAMMOMAT 3000 Nova only)
Check the "Flying Wing" locking mechanism (wing basic table). Swivel "wing 1"
("wing 1" is the table with the stereo connector) toward "wing 2" and back
(6/Fig.1/p.13).
SIM Level
Verify that the system is level with respect to the floor. Check the level screws
(4/Fig.1/p.13) with a water level and, if available, the floor mounting (optional).