6.2 Checks
SIE Emergency STOP
Press Emergency STOP, see chapter "Compression and system movements"
(4/Fig.1/p.20).
- Are the compression, rotation and vertical travel of the lifting carriage blocked?
Release the Emergency STOP.
PMF Indicators
Check the LEDs on the operating console by starting the panel test via the service PC.
PMP Error memory
Read out the error memory and the exposure counter with the service PC.
PMP Record the error memory.
PMP Delete the error memory.
Enter the data from the exposure counter and the error memory in the protocol and de-
lete the error memory only.
PMF Auxiliary voltages
Use the voltmeter to measure the auxiliary voltages according to circuit diagram
X041E, sheet 2-12/13 (test points on D801, D802 and D704).
System OFF
PMF Cassette locking
Check that the cassettes being used lock in properly to the object tables.
PMF OPDOSE
Check the selection of various anode/filter combinations and OPDOSE according to the
Installation Start-Up Instructions, chapter "Start-up and functional test of the ION-
TOMAT" or "Calibrating and adjusting the AEC".
PMF Stereo unit (Mammomat 3000 Nova only)
If attached, check the stereotactic biopsy option according to the supplement to the In-
structions for Use.
PMA UIs and SPEED info
Check whether all relevant UIs and SPEED info has been completed.