32 Miscellaneous
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02 Siemens AG
02.06 CS PS SP
Page 32 of 34
Medical Solutions
6.2 Checks
SIE Emergency STOP
• Press Emergency STOP, see chapter "Compression and system movements"
(4/Fig.1/p.20).
- Are the compression, rotation and vertical travel of the lifting carriage blocked?
• Release the Emergency STOP.
PMF Indicators
• Check the LEDs on the operating console by starting the panel test via the service PC.
PMP Error memory
• Read out the error memory and the exposure counter with the service PC.
PMP Record the error memory.
PMP Delete the error memory.
• Enter the data from the exposure counter and the error memory in the protocol and de-
lete the error memory only.
PMF Auxiliary voltages
• Use the voltmeter to measure the auxiliary voltages according to circuit diagram
X041E, sheet 2-12/13 (test points on D801, D802 and D704).
• System OFF
PMF Cassette locking
• Check that the cassettes being used lock in properly to the object tables.
PMF OPDOSE
• Check the selection of various anode/filter combinations and OPDOSE according to the
Installation Start-Up Instructions, chapter "Start-up and functional test of the ION-
TOMAT" or "Calibrating and adjusting the AEC".
PMF Stereo unit (Mammomat 3000 Nova only)
• If attached, check the stereotactic biopsy option according to the supplement to the In-
structions for Use.
PMA UIs and SPEED info
• Check whether all relevant UIs and SPEED info has been completed.