Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Compression and system movements 21
Page 21 of 34
Medical Solutions
4.2 Testing the "Decompression" function
• Attach the spring scale as shown in (1/Fig.2/p.21).
Tolerance: ±1 kp.
4.2.1 Switching off
SIE Safety switch
• Does decompression switch off at a counter pressure of approximately 5 kp?
¹ Because of the measurement location, this value corresponds to a value of
approximately 3 kp at the compression plate.
4.2.2 Travel
PMF Compression travel
• Move the compression plate up and down. Is the movement smooth and noiseless
through the entire range?
4.2.3 Applying oil and grease
PMA Apply oil and grease to the compression unit
Fig. 2: Compression unit with covers Fig. 3: Compression unit without covers