Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Column stand 17
Page 17 of 34
Medical Solutions
3-
3 Column stand
3.1 Checks
SIM Steel ropes
• Do the steel ropes (balancing and flap), the drive belt, the suspension, or the pulleys
have any damage (e.g. fraying or splitting) or show signs of wear?
PMA Oil and grease
• If they are OK, lightly oil the wire cable, grease the rails, and then remove the protective
strips.
SIE Limit switches
• System ON. Move the lifting carriage up and down several times, the lifting carriage
should be in an upright (0°) position.
NOTE
It is normal to hear some noise from the rotation of the radiation
filter in the collimator when switching the system ON.
• Do the limit switches stop carriage travel at the top and at the bottom?
• Activate safety switch S882. System movements must be blocked.
PMF Vertical travel
• Is the vertical travel smooth and noiseless?
• Switch OFF the system and attach the protective strips once again.
SIM Safety catch and rotation safety catch
• Do the safety catch (1/Fig.1/p.17) and rotation safety catch (2/Fig.1/p.17) have any
visible damage? Are the springs in good condition?
Fig. 1: Safety catch