Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Column stand 17
Page 17 of 34
Medical Solutions
3-
3 Column stand
3.1 Checks
SIM Steel ropes
Do the steel ropes (balancing and flap), the drive belt, the suspension, or the pulleys
have any damage (e.g. fraying or splitting) or show signs of wear?
PMA Oil and grease
If they are OK, lightly oil the wire cable, grease the rails, and then remove the protective
strips.
SIE Limit switches
System ON. Move the lifting carriage up and down several times, the lifting carriage
should be in an upright (0°) position.
NOTE
It is normal to hear some noise from the rotation of the radiation
filter in the collimator when switching the system ON.
Do the limit switches stop carriage travel at the top and at the bottom?
Activate safety switch S882. System movements must be blocked.
PMF Vertical travel
Is the vertical travel smooth and noiseless?
Switch OFF the system and attach the protective strips once again.
SIM Safety catch and rotation safety catch
Do the safety catch (1/Fig.1/p.17) and rotation safety catch (2/Fig.1/p.17) have any
visible damage? Are the springs in good condition?
Fig. 1: Safety catch