Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Miscellaneous 33
Page 33 of 34
Medical Solutions
6.3 Final tests
Remove the measurement devices.
PMF Operating problems
Were there any operating problems during the checks?
PMP Covers
Remove the protective strips and attach all removed covers. Check that all covers are
completely and firmly attached, including the cable duct.
SIE Protective conductor test
Perform the protective conductor test according to ARTD-002.731.17....
PMP Cleaning/damaged paint
Clean the unit using the materials recommended in the Instructions for Use and repair
any damaged paint.
System ON.
QSQ Final test exposure
Take one final test exposure.