24 Compression and system movements
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02 Siemens AG
02.06 CS PS SP
Page 24 of 34
Medical Solutions
4.4 Checking miscellaneous system movements
4.4.1 Checking miscellaneous system movements
SIE Switch off
• Check whether the motorized rotation movements shut off when a preset angle is
reached. The angle (e.g. 90°) is preset with the potentiometer, (3/Fig.1/p.13).
• Check the automatic switch-off function when passing the 0° position as well.
PMF Rotation movements
• Is the rotation movement smooth and noiseless through the entire range?
• Is the display correct (7/Fig.1/p.13)?
SIM Blocking rotation and vertical travel
• Rotation and vertical travel must be blocked at a compression force of > 3 kp.