Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Miscellaneous 31
Page 31 of 34
Medical Solutions
6-
6 Miscellaneous
6.1 Blocking exposure release
PMF Blocking
Exposure release must be blocked under the following conditions
(indicated on the control console):
- No cassette is inserted.
- Cassette was not changed after the last exposure.
- Table is not attached or not locked in place.
- Wing 1 or wing 2 is not locked in place. (Mammomat 3000 Nova only).
- For stereo only: no external collimator for beam limitation in the collimator.
(Mammomat 3000 Nova only).
- The two exposure release switches were not pressed simultaneously.
- No external collimator for 18 x 24 or magnification. (Mammomat 1000 only).