26 Test exposures
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02 Siemens AG
02.06 CS PS SP
Page 26 of 34
Medical Solutions
5.1 Evaluating test values
PMF Anode acceleration
Listen to the sound generated by the accelerating anode. Is the sound normal?
PMF kV and mA
Do the measured kV and mA values agree with the values selected?
PMF mAs selection
Do the measured mAs values agree with the values selected?
PMF mAs value displayed
Does the measured mAs value agree with the displayed mAs value (for AEC)?
PMF Grid voltage
Attach the magnification table. Measure the grid voltage (small focus) on tube test point
G against test point H
3
. Does the measured value agree with the value in the test proto-
col?
Test both focal spots (if available).
PMF Signal lamp
Does the radiation-ON indicator light up?