Siemens AG SPB7-230.101.02.05.02 MAMMOMAT 1000/3000 Nova
02.06 CS PS SP
Test exposures 25
Page 25 of 34
Medical Solutions
5-
5 Test exposures
System OFF
Connect the following measurement devices:
- Service PC to D702.X990
- Oscilloscope to MP kVactual.X711 (1V 5kV) and mAactual.X707,
OVA.X708 (1V 40mA)
- mAs meter to D710 (remove X3 - X4 jumpers)
- Multimeter at MP.BIAS on D706 (measurement range 200V=)
System ON
X
Test the kV/mA control loop according to the Installation & Start-Up Instructions, chap-
ter "Checks with high voltage".