22 Compression and system movements
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02 Siemens AG
02.06 CS PS SP
Page 22 of 34
Medical Solutions
• System OFF
Remove the upper and lower covers (snap closures, (2/Fig.2/p.21) and pot knob
(3/Fig.2/p.21).
• After removing the "small" Allen screws (two above and two below), the right cover
(4/Fig.2/p.21) of the compression unit can be removed.
• Grease the rails (2/Fig.3/p.21) lightly and check the belt (1/Fig.3/p.21) and string
(3/Fig.3/p.21) for damage/wear as well as for tension.
• Reinstall the covers on the compression unit.