Service and maintenance
10.1 Error and diagnostics messages
SIMATIC MV500
Operating Instructions, 05/2019, C79000-G8976-C494-02
177
Error
message
Value
Onboard
PROFINET
Value
TCP/IP
and
RS232
Value
Ident
profile,
STATUS
byte
Description
Corrective measures
Transfer error
archiving/
MMI
W#16#5007 07 E1FE04 The send buffer is full be-
cause the data cannot be
sent in an adequately short
time or the server is not
confirming receipt of the
data in time.
Reduce the trigger frequency or
reduce the amount of diagnostic
data to be archived. Process the
results on the server side faster.
A high load on the network to
which the optical reader is con-
nected can also block the trans-
fer of the data. If necessary,
check the network connection.
Transfer error
PROFINET
IO
W#16#5007 07 E1FE04 The send buffer is full be-
cause the data cannot be
queried in an adequately
short time.
Reduce the trigger frequency or
process the results faster. If
necessary, change the update
time in the PROFINET configu-
ration.
Transfer error
CM/
PROFINET
IO
W#16#5007 07 E1FE04 The send buffer is full be-
cause the data cannot be
queried in an adequately
short time.
Reduce the trigger frequency or
process the results more quickly
on the CPU. The transmission
speed of the communication
module connection can possibly
be increased.
To cover short-term peak loads,
the size of the image buffer in
the WBM can be increased un-
der"Program > Image acquisition
> Mode".
Program
requires
EF lens
W#16#5011 17 E1FE06 An EF lens is configured in
the program but it is not
connected.
Reconnect the EF lens or
change the parameter assign-
ment of the program.
Save pro-
gram
W#16#500A 10 E6FE04 The program could not be
created.
Adapt the parameter assign-
ment, the placement of test
objects in the image or the im-
age quality.
Match error W#16#500B 11 E1FE07 The program could not be
created due to faulty match
settings.
Adapt the match settings or use
a suitable test object.
Timeout
during saving
W#16#500D 13 E6FE04 The "Save program" process
was canceled because it
took too long.
Check the program settings and
start the "Save program" pro-
cess via the WBM, if necessary.
TCP commu-
nication
W#16#500E 14 E1FE01 Connection error in commu-
nication over a TCP connec-
tion or when using an
Ethernet RS-232 converter:
There is no TCP connection
or the data transfer cannot
be completed within the set
time limit.
Make sure that the connected
TCP partner is correctly config-
ured and ready to receive.