Preparations 2
• Switch the SIREMOBIL off and disconnect the power plug.
• Secure the system against being switched back on.
• For safety against tilting, support the system on the operating side with wood or
something similar.
Board D3 2
• Unscrew the plexiglas cover of the board D3.
• Pull off all connectors except for D3/X2 (Fig.1).
• Unscrew the protective conductor from the board.
Board D1 2
• Pull all connectors off from the board.
• Unscrew the grounding strap from the board.
• Unscrew all 6 fastening nuts (1/Fig.2)
• Remove the board carefully and place it on a surface suitable for ESD.
• Remove the spacer bushes from the 6 screws.
• Open the strain relief of the ribbon cable X 4 (2/Fig.2).
Fig. 1 Fig. 2
The board D3 does not have to be removed.
1
2
NOTE
Removing the lifting column 2