SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 10 of 10 Siemens AG
System Manual Rev. 01 07.05 CS PS 24 Medical Solutions
3 - 10 Installing the lifting column
Functional check 3
Move the lifting column out. Press the emergency stop switch approximately in the
middle of the vertical lift.
- It must be no longer possible to move the lifting column.
Unlock the emergency stop switch again.
- The lifting column can be moved again.
Move the lifting column into the top end position.
- The lifting column motor must switch off automatically at the top point of the lift.
- No squeaking noises should occur during the movement of the lifting column.
Move the lifting column in completely.
- The lifting column motor must switch off before the lowest end position is reached and
an acoustic warning signal must be audible.
- No squeaking noises should occur during the movement of the lifting column.
Briefly release the button for lowering the lifting column and then press it again.
- The lifting column can be moved into the lowest position. An acoustic warning signal
sounds during the movement into the lowest end position.
Protective conductor test 3
Perform the protective conductor test according to ARTD-002-731.17....
Observe the instructions for the protective conductor test in the first chapter.
¹ The protective conductor resistance must not exceed 0.2 ohms.