EasyManua.ls Logo

Siemens SIREMOBIL Compact L - Adjustments; Programs; Generator, kV offset adjustment

Siemens SIREMOBIL Compact L
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installing the lifting column 3 - 7
Siemens AG n.a. SPR2-130.841.02 Page 7 of 10 SIREMOBIL Compact L
Medical Solutions Rev. 01 07.05 CS PS 24 System Manual
Installing the switch-mode power supply M 14 3
Install the base plate.
Install the switched-mode power supply with the plug-in connections downwards.
Plug in the connections.
Screw the protective conductor onto the base plate of the line filter and place the contact
washer under the cable lug.
Installing the stabilization bracket 3
Fasten the stabilization bracket of the monitor cable plug again.
Fasten the protective conductor on the unit floor and on the chassis, and place the
contact washers under the cable lugs.
Installing the boards D1 and D2 3
Fasten the heat sink on the lifting column operating side (Fig. 13).
Fasten 6 spacer screws on the heat sink (Fig. 13).
Apply heat conduction paste on the contact surface for the the heat sink.
Fit the board D2 on the spacer screws and fasten it to the heat sink (Fig. 13).
Plug in the connectors on D2 board.
Fasten the D2 board with the spacers (Fig. 13).
Plug the intermediate plate back on.
Fit the spacer tubes on the 6 screws.

Other manuals for Siemens SIREMOBIL Compact L