SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 8 of 10 Siemens AG
System Manual Rev. 01 07.05 CS PS 24 Medical Solutions
3 - 8 Installing the lifting column
• Fit the board D1 carefully on the 6 screws and fasten with washers and nuts (Fig. 14).
• Plug the plugs on the corresponding connectors of D1 board.
• Fasten the grounding strap on the D1 board as well as on the intermediate plate, on the
heat sink and on the base plate of the line filter.
When tightening the screws on the D1 board make sure that the grounding strap has no
contact with the components in its vicinity.
• Connect all shields of the connected cables to ground with cable clamps.
• Fasten the C-arm cable (1/Fig. 14) and plug in the connectors.
• Draw the Diamentor cable through to the left behind the lifting column.
• Fasten the bracket (2/Fig. 15) back of the lifting column.
Fig. 14 Fig. 15
Fig. 16
2
1