SIREMOBIL Compact L n.a. SPR2-130.841.02 Page 4 of 10 Siemens AG
System Manual Rev. 01 07.05 CS PS 24 Medical Solutions
3 - 4 Installing the lifting column
Installing the new cover hood with control board 3
Fit the cover hood over the lifting column.
Guide the ribbon cable and the protective conductor of the control board through the
recess (1/Fig. 7).
Place the cover hood in position.
Fasten the cover hood to the lifting column (Fig. 8) with 4 screws and to the steering lever
side from below also with 4 screws; the two long screws are intended for the inside.
Refit the strain relief (2/Fig. 7) of the control panel cable.
Screw the protective conductor (3/Fig. 7) back onto the chassis (on the operating side
from the left below).
Refit the small rectangular cover panel underneath the steering lever (2 paneling
screws).
Fig. 7 Fig. 8
Fig. 9 Fig. 10